Analisis Linearitas Keluaran Radiasi pada X-Ray Mobile dengan Menggunakan Piranha
Abstract
The research was conducted on the linearity of the output radiation in the mobile X-ray by using Piranha. This study aimed to measure the radiation dose exposure, to determine the relationship between the increase in the voltage of the radiation dose exposure and increase the tube current to the radiation dose exposure, as well as to analyze the radiation output and linearity of the output radiation. Data collection was conducted at the Dr. Tadjuddin Chalid Hospital Makassar in radiological installation. In this study the variables that measured the radiation dose and exposure variables that change the tube current and voltage. The data obtained and analyzed to calculate the value of the output radiation and radiation output linearity in the X- ray mobile. The analysis showed that the X-ray mobile with GE brand that are in radiological installation Dr. Tadjuddin Chalid Hospital has good linearity.
Downloads
References
Anugerah Firmansyah, 2012, Analisis Kebocoran Radiasi Pada Phywe X-Ray Unit Dengan Surveymeter, Skripsi,.
Aulya Rahayu, 2011, Perbandingan Karakteristik Keluaran Antara Pesawat Sinar-x Toshiba Model DRX-1824B dan Toshiba Model DRX-1603B. BATAN, Layanan Compliance Test Pesawat Sinar-x, http://www.batan.go.id/ptkmr/labptkmr/compliance_test.html (04 Mei 2013).
Departemen Kesehatan RI Direktotat Jendral Pelayanan Medik, 2001, Pedoman Pengujian dan Kalibrasi Alat Kesehatan. Jakarta : DEPKES RI.
Donohue, Daniel P., 2000, An Analysis of Radiographic Quality. New York : PRO-ED, Incorporated.
Dwi Oktavina Winata, 2011, Koefisien Backscatter Faktor Sinar-x Diagnostik dalam Rentang RQR (Radiation Qualities In Radiodiagnostik).
Healt Departmen of Western Australia, 2008, Workbook 1 Mobile Radiographic Equipment. 2nd edition(I). Australia: Healt Departmen of Western Australia. http://www.medgadget.com/2011/03/mobilett_mira_mobile_wireless_xray_from_siemens.html (update: 4/5/2013).
Khan, Faiz M., 2003, The Physics of Radiation Therapy. 2nd edition.
Philadelphia, USA: Lippincott Williams & Wilkins.
Ruslan Hani, Ahmadi dan Handoko Riwidikdo., 2009, Fisika Kesehatan. Jogjakarta: Mitra Cendekia Press.
Surat Keputusan Kepala BAPETEN Nomor 01-P/Ka-BAPETEN/ I-03 tentang Pedoman Dosis Pasien Radiodiagnostik, Jakarta , 2003.
Authors who publish with this journal agree to the following terms:
1) Authors retain copyright and grant the journal right of first publication with the work simultaneously licensed under a Creative Commons Attribution License that allows others to share the work with an acknowledgement of the work's authorship and initial publication in this journal.
2) Authors are able to enter into separate, additional contractual arrangements for the non-exclusive distribution of the journal's published version of the work (e.g., post it to an institutional repository or publish it in a book), with an acknowledgement of its initial publication in this journal.
3)Authors are permitted and encouraged to post their work online (e.g., in institutional repositories or on their website) prior to and during the submission process, as it can lead to productive exchanges, as well as earlier and greater citation of published work (See The Effect of Open Access).